ProfiClave New Generation PCX

1000400-0
On request - please contact us!

Description

Dimensions (HxWxD): 50x56x58 cm (lid open: H = 83 cm)

Weight: 64 kg

Ideal working Volume in Media preparation Mode: 2,5 - 15,5 L (vessel)

Accuracy: ±0,5 °C (resolution: 0,1 °C)

Maximum sterilisation temperature: 122 °C

Power consumption: 3500 W

Operating voltage AC: 200 – 240 V 50/60 Hz

Fuse: 16 A

Water connection: 3/4"

Maximum process pressure: 1,8 bar

Safety valve: 2,0 bar

Place of use: Laboratory-use only Type III Grade water e.g. deionized or
distilled water

Air pollution level: Degree of pollution 2 (IEC EN 61010-1)

Over voltage category: II

Ambient temperature: 10…30 °C

Cooling water: 5…30 °C at 0,8 – 5 bar

Air filter: 0,2 um PTFE sterile filter

Traceability: CSV file via USB

Traceability: Maintenance menu with Timestamp and Initials 

Level of access: 3

Options